|
Basic
ECG/Pacemaker Timing
жалюзи . пошив штор . доставка цветов киев . корм для собак Last update: 17 October 2000
| The Basic Timing Intervals
|
|
|
Typically, electrocardiograms (ECG's) are performed with a paper speed of
25mm/sec. When this is the case, each "large" box represents an interval
of 200 milliseconds (mS)or 0.2 seconds. The "small" boxes or timing marks within
each 200 mS section, represent an interval of 40 mS or 0.04 seconds. One millisecond
is equal to 1/1000th of 1 second, therefore, 1000 mS is equal to one second. One
second of time, as displayed on the ECG strip, is equal to 5 large boxes.
Pacemaker
timing is always described in mS intervals. An interval is defined as the time
between two consecutive events. A pacemaker rate is the average of several intervals
calculated for one minute of time.
By applying the "60,000" rule, it is easy to convert seconds to milliseconds or
mS to seconds. 60,000 divided by the rate returns the millisecond interval. If
60,000 is divided by the mS interval, the appropriate rate is given. |
| |
|
| Basic Pacemaker Intervals |
|
| In order for a pacemaker output to occur, an escape interval must have timed
out. By measuring the spike to spike interval, in the example above, and dividing
the interval into 60,000 the basic programmed rate can be determined. The
rate/interval table below, lists some commonly encountered intervals, and may be
helpful. |
| RATE
|
INTERVAL
|
RATE
|
INTERVAL
|
RATE
|
INTERVAL
|
| 30 bpm |
2000 mS |
60 bpm |
1000 mS |
100 bpm |
600 mS |
| 50 bpm |
1200 mS |
70 bpm |
857 mS |
120 bpm |
500 mS |
|
Sensitivity
The Sensitivity setting is the parameter that allows the pacemaker to "see"
intrinsic events. The illustration shows a device programmed to 4 mV of sensitivity
(indicated by black line). The "R" waves that fall below the black line
cannot be sensed. Common practice requires a 2:1 safety margin for sensing; the 6 mV
"R wave" does not meet the criteria and, although sensed, I have color coded it
with an amber color as a warning to proceed with caution. Signals of 8 mV or greater
are well within the normal sensing range of this device. |
|
|
The Refractory Period
The Refractory Period is a brief period of time when the device is not allowed to look for
intrinsic events. Refractory periods are usually divided into two segments; the
absolute portions and the relative (or technical) portion. Nothing can be sensed
during the absolute segment. The relative refractory period allows sensing to occur;
however, the pacemaker does not react to the event by resetting an escape interval.
In the illustration at the left, the Alert period represents when "normal"
sensing is allowed to occur. |
Output
The electrical output of a pacemaker is determined by the amount of Voltage applied to the
heart and the amount of time the Voltage is applied. All implantable pacemakers are
constant Voltage output devices. The pacemaker output consists of the following
parts: |
|
| 1. |
The Voltage Output - this is the electrical "pressure" that is
applied to electrons to get them moving down the pacing lead wire |
| 2. . |
The Pulse Width - this is the time in milliseconds that the Voltage is
applied |
| 3. |
The Discharge Slope - this is dependent upon the resistance placed on the
flow of electrons and other electrical variables. |
|